Equipment and Calibration History

History for YSI EXO pH-ORP 24G102326

Rec No.DateLoc FromTitleLoc ToNotes
7148 2025-02-27 Lab none Calibration Lab none ORP (mV), Last Cal: 2024-10-22
EXO sn 23H107001
EXO/Sensor Firmware 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,227.0 mV,218.3 mV,227.0 mV,0.0 mV,22.018 °C
Note: Old Zobell solution (mixed 2022-09-23) only about 0.4mV different from fresh Zobell solution mixed today and stored in small brown bottle. Old Zobell solution stored in its original large brown bottle.
7147 2025-02-27 Lab none Calibration Lab none pH, Last cal: 2024-10-22
EXO sn 23H107001
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.01 pH,7.11 pH,7.01 pH,-18.6 mV,20.021 °C
Pt2,10.06 pH,10.18 pH,10.06 pH,-192.5 mV,19.981 °C
Pt3,4.00 pH,4.09 pH,4.00 pH,153.2 mV,20.023 °C
4-7Slope,4-7mVPerDecade,7-10Slope,7-10mVPerDecade
171.5 mV,56.98 mV,174.1 mV,57.09
Note: Using standards from 2023-12-04
7142 2025-02-18 Gilbert Tract pH ORP Removed Lab none Removed from the GT channel and brought back to the lab for routine calibration
6945 2024-10-23 Lab none Installed Gilbert Tract pH ORP Installed in the GT channel after calibration
7045 2024-10-22 Lab none Calibration Lab none pH, Last cal: 2024-07-25
EXO sn 23H107001
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.01 pH,7.03 pH,7.01 pH,-12.6 mV,21.535 °C
Pt2,10.06 pH,10.08 pH,10.06 pH,-186.6 mV,22.141 °C
Pt3,4.00 pH,4.01 pH,4.00 pH,160.2 mV,22.907 °C
4-7Slope,4-7mVPerDecade,7-10Slope,7-10mVPerDecade
172.8 mV,57.41 mV,174.1 mV,57.07
Note: Used opened standards from Dec 2023.
7044 2024-10-22 Lab none Calibration Lab none ORP (mV), Last Cal: 2024-07-25
EXO sn 23H107001
EXO/Sensor Firmware 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,226.7 mV,222.3 mV,226.7 mV,0.0 mV,23.160 °C
Note: Used Zobell solution (mixed 2022-09-23) in small brown plastic bottle.
6929 2024-10-17 Gilbert Tract pH ORP Removed Lab none Removed from GT channel and brought back to the lab for calibration
6830 2024-08-16 Lab none Installed Gilbert Tract pH ORP Installed with EXO3s in Gilbert Tract channel, co-located with the ADCP and channel miniDOT
7047 2024-07-25 Lab none Calibration Lab none pH, Last cal:
EXO sn 23H107001
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.01 pH,7.19 pH,7.01 pH,-10.9 mV,22.992 °C
Pt2,10.06 pH,10.18 pH,10.06 pH,-186.1 mV,22.810 °C
Pt3,4.00 pH,4.26 pH,4.00 pH,160.6 mV,22.822 °C
4-7Slope,4-7mVPerDecade,7-10Slope,7-10mVPerDecade
171.5 mV,56.98 mV,175.2 mV,57.43
Note: First calibration on a new sensor. Used opened standards from Dec 2023.
7046 2024-07-25 Lab none Calibration Lab none ORP (mV), Last Cal:
EXO sn 23H107001
EXO/Sensor Firmware 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,223.0 mV,228.6 mV,223.0 mV,0.0 mV,23.893 °C
Note: First calibration on a new sensor. Used Zobell solution (mixed 2022-09-23) in small brown plastic bottle.