Equipment and Calibration
View History
Equipment Entry
|
Documents
No documents found
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6962 | 2024-10-24 | Lab none | Installed | Hill Slough fDOM | Installed in the HS channel after calibration |
7039 | 2024-10-22 | Lab none | Calibration | Lab none | fDOM (QSU and RFU), Last cal: 2024-07-22 EXO sn 22C103416 EXO/Sensor Firmware: 1.0.86/3.0.0 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,0.10 QSU,0.00 QSU,0.00 QSU,23.760 °C Pt2,300.00 QSU,319.62 QSU,300.01 QSU,0.00 QSU,22.150 °C Notes: Calibration solution from 2024-07-21. First calibration on new sensor. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,-.02 RFU,0.00 RFU,0.00 RFU,23.208 °C Pt2,100.0 RFU,106.05 RFU,100.00 RFU,0.00 RFU,22.127 °C Notes: Cal solution using QSE solution from 2024-07-20. |
6935 | 2024-10-17 | Hill Slough fDOM | Removed | Lab none | Removed from the Hill Slough channel and brought back to the lab for routine calibration |
6777 | 2024-07-23 | none | Installed | Hill Slough fDOM | Installed in the Hill Slough channel after calibration. Sonde is in the air during low tide. |
6791 | 2024-07-22 | Lab none | Calibration | Lab none | fDOM (QSU and RFU), Last cal: EXO sn 22C103416 EXO/Sensor Firmware: 1.0.86/3.0.0 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,-1.64 QSU,0.00 QSU,0.00 QSU,22.291 °C Pt2,310.80 QSU,283.56 QSU,310.81 QSU,0.00 QSU,14.025 °C Notes: Calibration solution from 2024-07-21. First calibration on new sensor. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,-.53 RFU,0.00 RFU,0.00 RFU,22.076 °C Pt2,104.1 RFU,94.92 RFU,104.10 RFU,0.00 RFU,13.123 °C Notes: Calibration solution from 2024-07-21. First calibration on new sensor. |