Equipment and Calibration History

History for YSI EXO fDOM 24F104438

Rec No.DateLoc FromTitleLoc ToNotes
6962 2024-10-24 Lab none Installed Hill Slough fDOM Installed in the HS channel after calibration
7039 2024-10-22 Lab none Calibration Lab none fDOM (QSU and RFU), Last cal: 2024-07-22
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,0.10 QSU,0.00 QSU,0.00 QSU,23.760 °C
Pt2,300.00 QSU,319.62 QSU,300.01 QSU,0.00 QSU,22.150 °C
Notes: Calibration solution from 2024-07-21. First calibration on new sensor.

,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,-.02 RFU,0.00 RFU,0.00 RFU,23.208 °C
Pt2,100.0 RFU,106.05 RFU,100.00 RFU,0.00 RFU,22.127 °C
Notes: Cal solution using QSE solution from 2024-07-20.
6935 2024-10-17 Hill Slough fDOM Removed Lab none Removed from the Hill Slough channel and brought back to the lab for routine calibration
6777 2024-07-23 none Installed Hill Slough fDOM Installed in the Hill Slough channel after calibration. Sonde is in the air during low tide.
6791 2024-07-22 Lab none Calibration Lab none fDOM (QSU and RFU), Last cal:
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,-1.64 QSU,0.00 QSU,0.00 QSU,22.291 °C
Pt2,310.80 QSU,283.56 QSU,310.81 QSU,0.00 QSU,14.025 °C
Notes: Calibration solution from 2024-07-21. First calibration on new sensor.

,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,-.53 RFU,0.00 RFU,0.00 RFU,22.076 °C
Pt2,104.1 RFU,94.92 RFU,104.10 RFU,0.00 RFU,13.123 °C
Notes: Calibration solution from 2024-07-21. First calibration on new sensor.