History for YSI EXO Turbidity 24F102410
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
7145 | 2025-02-27 | Lab none | Calibration | Lab none | Turbidity (FNU), Last cal: 2024-10-22 EXO sn 23H107001 EXO/Sensor Firmware: 1.0.86/3.0.5 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 FNU,-0.04 FNU,0.00 FNU,0.00 FNU,23.311 °C Pt2,124.00 FNU,114.35 FNU,124.01 FNU,0.00 FNU,23.413 °C Pt3,1010.00 FNU, 931.74 FNU, 1009.99 FNU, 0.00 FNU, 23.782 °C Note: Much larger offset for 1010 FNU point that we have seen before, brand new cal solutions |
7141 | 2025-02-18 | Gilbert Tract Turbidity | Removed | Lab none | Removed from the GT channel and brought back to the lab for routine calibration |
6942 | 2024-10-23 | Lab none | Installed | Gilbert Tract Turbidity | Installed in the GT channel after calibration |
7042 | 2024-10-22 | Lab none | Calibration | Lab none | Turbidity (FNU), Last cal: 2024-07-25 EXO sn 23H107001 EXO/Sensor Firmware: 1.0.86/3.0.5 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 FNU,-0.06 FNU,0.00 FNU,0.00 FNU,23.324 °C Pt2,124.00 FNU,123.12 FNU,124.00 FNU,0.00 FNU,22.559 °C Pt3,1010.00 FNU, 1004.19 FNU, 1010.01 FNU, 0.00 FNU, 22.612 °C Note: |
6926 | 2024-10-17 | Gilbert Tract Turbidity | Removed | Lab none | Removed from GT channel and brought back to the lab for calibration |
6826 | 2024-08-16 | Lab none | Installed | Gilbert Tract Turbidity | Installed with EXO3s in Gilbert Tract channel, co-located with the ADCP and channel miniDOT |
7049 | 2024-07-25 | Lab none | Calibration | Lab none | Turbidity (FNU), Last cal: EXO sn 23H107001 EXO/Sensor Firmware: 1.0.86/3.0.5 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 FNU,-0.04 FNU,0.00 FNU,0.00 FNU,23.318 °C Pt2,124.00 FNU,131.81 FNU,124.00 FNU,0.00 FNU,23.157 °C Pt3,1010.00 FNU,983.43 FNU,1010.01 FNU,0.00 FNU,23.294 °C Note: First calibration on new sensor. |