Equipment and Calibration History
History for YSI EXO Wiped Cond/Temp 24F100050
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
7146 | 2025-02-27 | Lab none | Calibration | Lab none | Sp Cond (uS/cm),Last cal: 2024-10-22 EXO sn 23H107001 EXO/Sensor Firmware 1.0.86/3.0.5 CellConstant 0.47 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,10000.0 µS/cm, 10035.5 µS/cm, 10000.0 µS/cm, 0.0 µS/cm,17.473 °C Note: Old 10,000 uS/cm solution from 2022 had some dirt in it and read about 200uS/cm higher than the fresh used for the calibration. We discarded the old solution. |
7140 | 2025-02-18 | Gilbert Tract Conductivity Sensor Channel | Removed | Lab none | Removed from the GT channel and brought back to the lab for routine calibration |
6944 | 2024-10-23 | Lab none | Installed | Gilbert Tract Conductivity Sensor Channel | Installed in the GT channel after calibration |
7043 | 2024-10-22 | Lab none | Calibration | Lab none | Sp Cond (uS/cm),Last cal: 2024-07-25 EXO sn 23H107001 EXO/Sensor Firmware 1.0.86/3.0.5 CellConstant 0.47 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,5000.0 µS/cm, 4991.9 µS/cm,5000.0 µS/cm, 0.0 µS/cm,22.658 °C Note: Used USGS standard from Fondriest that expired Aug 2023. It was stored in fridge w/ cap taped and I brought it close to room temperature before calibration. |
6928 | 2024-10-17 | Gilbert Tract Conductivity Sensor Channel | Removed | Lab none | Removed from GT channel and brought back to the lab for calibration |
6828 | 2024-08-16 | Lab none | Installed | Gilbert Tract Conductivity Sensor Channel | Installed with EXO3s in Gilbert Tract channel, co-located with the ADCP and channel miniDOT |
7048 | 2024-07-25 | Lab none | Calibration | Lab none | Sp Cond (uS/cm),Last cal: EXO sn 23H107001 EXO/Sensor Firmware 1.0.86/3.0.5 CellConstant 0.47 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,5000.0 µS/cm,4954.6 µS/cm,5000.0 µS/cm,0.0 µS/cm,22.816 °C Note: First calibration on new sensor. Used USGS standard from Fondriest that expired Aug 2023. It was stored in fridge w/ cap taped and I brought it close to room temperature before calibration. |