Rec No. | Date | Loc From | Title | Loc To | Notes |
6729 |
2024-05-16 |
Lab none |
Installed |
Hill Slough Water Level |
Installed at the HS channel to replace CTD sn H21G200002433 that was removed previously. Tied to the same hook as the eosGP and I tried to match the eosGP height. |
6664 |
2024-04-18 |
Lab none |
Tested |
Lab none |
I cut the chewed portion of the cable off and stripped the wires. I connected it to a CR1000 and changed its address from 0 to 6 to match the program I had. It seemed to work fine. No obvious water in the vent tub but the broken part of the cable was under water. We will have to put a desiccant tube on this because the fancy Teflon filter can't be reattached easily. |
6661 |
2024-04-16 |
Gilbert Tract Water Level |
Removed |
Lab none |
Removed from GT and brought back to the lab. Its cable was chewed and totally broken in at least 2 places. |
6047 |
2022-10-19 |
Lab none |
Installed |
Gilbert Tract Water Level |
Installed at GT off of the boardwalk where it is co-located with the eosGP and Campbell conductivity sensor. When the staff gauge read 21cm, the bottom of the sensors was 19cm below the water. |