Equipment and Calibration

View History

Equipment Entry

Instrument:Soil Water Potential
Manufacture:METER Group
Model No:TEROS21
Serial No:G20007096
Tag No:
Acquisition Date:2022-03-21
Project:Ameriflux
Location on 2025-05-15:Tonzi,
Notes: Given to us by Ameriflux four Year-of-Water emphasis - to be installed at Tonzi

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
7236 2025-05-15 Lab none Installed Tonzi Soil Water Potential C Installed on the OSU SP1 system at about -10cm in its previous location. The current program isn't reading it but I could read it directly from the terminal - not sure what's going on.
7202 2025-04-24 Lab none Tested Lab none Tested on a CR1000 in the lab - seems to be okay. Terminal SDI-12 interface worked okay and the sensor responded to water. Try reinstalling it in the field.
7201 2024-12-03 Tonzi Soil Water Potential C Removed Lab none Removed from near OSU SP1 because of bad readings. No obvious damage.
6796 2024-05-15 Lab none Installed Tonzi Soil Water Potential C Reinstalled in close to the original location at about -10cm after repairing the cow chewed cable in the lab
6723 2024-05-08 Tonzi Soil Water Potential C Removed Lab none Removed from the OSU SP1 system at Tonzi for testing. It wasn't giving data. Cows might have damaged its cable.
5930 2022-06-07 Lab none Installed Tonzi Soil Water Potential C Installed near the OSU-SP1 soil sensors at -10cm, vertical orientation