Equipment and Calibration History
History for METER Group TEROS21 G20007096
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
7236 | 2025-05-15 | Lab none | Installed | Tonzi Soil Water Potential C | Installed on the OSU SP1 system at about -10cm in its previous location. The current program isn't reading it but I could read it directly from the terminal - not sure what's going on. |
7202 | 2025-04-24 | Lab none | Tested | Lab none | Tested on a CR1000 in the lab - seems to be okay. Terminal SDI-12 interface worked okay and the sensor responded to water. Try reinstalling it in the field. |
7201 | 2024-12-03 | Tonzi Soil Water Potential C | Removed | Lab none | Removed from near OSU SP1 because of bad readings. No obvious damage. |
6796 | 2024-05-15 | Lab none | Installed | Tonzi Soil Water Potential C | Reinstalled in close to the original location at about -10cm after repairing the cow chewed cable in the lab |
6723 | 2024-05-08 | Tonzi Soil Water Potential C | Removed | Lab none | Removed from the OSU SP1 system at Tonzi for testing. It wasn't giving data. Cows might have damaged its cable. |
5930 | 2022-06-07 | Lab none | Installed | Tonzi Soil Water Potential C | Installed near the OSU-SP1 soil sensors at -10cm, vertical orientation |