Equipment and Calibration
View History
Equipment Entry
|
Documents
No documents found
Report: HMP intercomparison - BA 11/2021 - 2021-11-04
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6557 | 2024-01-25 | Factory none | Disposed | Disposed none | Campbell's evaluation: Inspected. Cap filter is dirty and sticker labels started to peel off suggesting sensor has been deployed in a very humid environment. After tests sensor showed problems to provide correct RH and Temp data. Unfortunately sensor can't be repaired or calibrated. Warranty expired o 08-27-2022. NOTE: We suggest to acquire a new unit such as the HygroVUE10, which offers advantages worth look up, as a replacement for the damaged one. Dead unit will be returned so that I can take it apart! |
6556 | 2024-01-17 | Lab none | Shipped | Factory none | Sent to Campbell on RMA 37957. RH and AirT are reading zero. |
6485 | 2023-12-08 | Mayberry RH & Temp | Removed | Lab none | Removed from the Mayberry tower and brought back to the lab for testing. It was reading -79 deg C and 0% RH. |
5673 | 2021-11-23 | Lab none | Installed | Mayberry RH & Temp | Installed on the MB tower to replace HMP45 sn U2120013 which was drifting. It was installed with a new shield and aspirated with a 12V fan. The cable on this sensor says "sn T2820183", which is different than the actual sn of the sensor, so I crossed out sn on the cable to avoid confusion. |
5585 | 2021-11-18 | Cal Kit RH & Temp | Removed | Lab none | Removed from BA tower at the conclusion of the HMP155 intercomparison. |
5565 | 2021-11-05 | Bouldin Alfalfa-ADEX RH & Temp Bouncer Upper | Moved | Cal Kit RH & Temp | Removed from bouncer tower (upper position) at the conclusion of this year's advection experiment and installed on the main tower for an HMP intercomparison |
5449 | 2021-09-03 | Lab none | Installed | Bouldin Alfalfa none | Installed on the BA bouncer tower for the advection experiment in the upper T/RH location. |