Equipment and Calibration History

History for Vaisala HMP155 T2820180

Rec No.DateLoc FromTitleLoc ToNotes
6557 2024-01-25 Factory none Disposed Disposed none Campbell's evaluation:
Inspected. Cap filter is dirty and sticker labels started to peel off suggesting sensor has been deployed in a very humid environment. After tests sensor showed problems to provide correct RH and Temp data. Unfortunately sensor can't be repaired or calibrated. Warranty expired o 08-27-2022. NOTE: We suggest to acquire a new unit such as the HygroVUE10, which offers advantages worth look up, as a replacement for the damaged one.

Dead unit will be returned so that I can take it apart!
6556 2024-01-17 Lab none Shipped Factory none Sent to Campbell on RMA 37957. RH and AirT are reading zero.
6485 2023-12-08 Mayberry RH & Temp Removed Lab none Removed from the Mayberry tower and brought back to the lab for testing. It was reading -79 deg C and 0% RH.
5673 2021-11-23 Lab none Installed Mayberry RH & Temp Installed on the MB tower to replace HMP45 sn U2120013 which was drifting. It was installed with a new shield and aspirated with a 12V fan. The cable on this sensor says "sn T2820183", which is different than the actual sn of the sensor, so I crossed out sn on the cable to avoid confusion.
5585 2021-11-18 Cal Kit RH & Temp Removed Lab none Removed from BA tower at the conclusion of the HMP155 intercomparison.
5565 2021-11-05 Bouldin Alfalfa-ADEX RH & Temp Bouncer Upper Moved Cal Kit RH & Temp Removed from bouncer tower (upper position) at the conclusion of this year's advection experiment and installed on the main tower for an HMP intercomparison
5449 2021-09-03 Lab none Installed Bouldin Alfalfa none Installed on the BA bouncer tower for the advection experiment in the upper T/RH location.