Rec No. | Date | Loc From | Title | Loc To | Notes |
5476 |
2021-10-06 |
Tonzi |
Calibration |
Tonzi none |
Tested with our new method using the peristaltic pump and 237ml of water which should result in 50 tips. 1/2 turn of the screws (CW+, CCW-) should adjust the number of tips by 2-3% (1-1.5 tips). Got 52 tips – I turned the screws one turn CCW. |
4143 |
2019-11-07 |
Tonzi |
Calibration |
Tonzi |
Plug tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket which should result in 100tips. Got 103 tips and did not adjust the bucket. |
3738 |
2018-10-17 |
Tonzi |
Calibration |
Tonzi |
Got 134 of 100 tips - adjusted screws 4 turns CCW |
3637 |
2012-01-09 |
Lab none |
Installed |
Tonzi |
Installed on the ground at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ. |