Equipment and Calibration

View History

Equipment Entry

Instrument:Precipitation
Manufacture:Texas Electronics
Model No:TE525
Serial No:48102-611
Tag No:
Acquisition Date:2018-06-26
Project:OSU AirMOSS
Location on 2021-10-06:Tonzi,
Notes: Acquired from the OSU AirMOSS project, Location SP03-23

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
5477 2021-10-06 Tonzi Calibration Tonzi none Tested with our new method using the peristaltic pump and 237ml of water which should result in 50 tips. 1/2 turn of the screws (CW+, CCW-) should adjust the number of tips by 2-3% (1-1.5 tips). Got 50 tips – No adjustment
4144 2019-11-07 Tonzi Calibration Tonzi Plug tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket which should result in 100tips. Got 103 tips and did not adjust the bucket.
3617 2012-01-09 Lab none Installed Tonzi Installed on the ground at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ.