Equipment and Calibration History
History for Global Water WL16U 57278
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5084 | 2018-05-11 | Lab none | Installed | Tonzi Water Level Road | Installed at the Road Well to measure water level at 5min intervals storing 30min averages |
5087 | 2018-05-08 | Lab none | Repair | Lab none | I stripped off the old rotting insulation and then slid on a long continuous piece of heat shrink. I shrunk the whole length of the cable. I disassembled the data logging enclosure and kept only the data logger board and USB connector in the end cap. I cut off the twin 9V battery plugs and will power the sensor from solar connected rechargeable battery. |
5086 | 2018-03-27 | Tonzi Water Level Floor | Removed | Lab none | Removed from the Floor Well. The cable insulation is rotting and needs to be replaced. Replaced by the new METER water level sensor a couple of weeks ago. |