Rec No. | Date | Loc From | Title | Loc To | Notes |
7122 |
2025-01-30 |
Lab none |
Installed |
Bouldin Alfalfa none |
Installed at BA to collect raw and 30min data from the LI-720 prototype |
7105 |
2025-01-27 |
Lab none |
Upgrade |
Lab none |
I updated the OS from v7.02 to v14.01. The first attempt over the CSIO port failed but the second attempt via the USB port worked fine. Seems like a lot of encryption/security changes in v14. |
6463 |
2023-11-07 |
Loaned none |
Returned |
Lab none |
Returned from Yuankun Xu (Roland Burgmann Lab) and Yifang Cheng at the conclusion of their field project at the Slumgullion Landslide site in Colorado. |
6380 |
2023-08-22 |
Lab none |
Loaned |
Loaned none |
Loaned to Yuankun Xu (Roland Burgmann Lab) and Yifang Cheng (Berkeley Seismo Lab) for use at Slumgullion Landslide site in Colorado for about one month mid-Sept to mid-Oct. Loaned with sonics sn 152603 and 153804 and CR6 sn 2115. |
5857 |
2022-04-25 |
Lab none |
Installed |
Bouldin Alfalfa-ADEX Data Logger Bouncer Flux |
Installed on the Bouncer tower to collect 20Hz data from the lower sonic and 10sec data from the two RH/Temp sensors for the advection experiment at BA |
5812 |
2022-04-22 |
Lab none |
Maintenance |
Lab none |
The internal lithium battery was low ~0.9V. I replaced it with a "slightly used" Lithium AA battery that was reading 3.2V on the voltmeter. |
3312 |
2018-01-31 |
Factory |
Returned |
Lab none |
Returned from Campbell where it was evaluated, updated and returned free of charge. |
3311 |
2018-01-22 |
Factory |
Maintenance |
Factory |
Campbell says: Tested unit on bench, unit functioned upon receipt. No problems found, but unit is running OS Rev3. This Rev is suspected to cause random freezes that require disconnection of the lithium battery. The issue has been resolved in the latest CR6 OSes. Updated the logger to latest Rev 7.02. Performed calibration, tested unit in the environmental chamber, and performed final bench test. Replaced the internal lithium battery and included a new USB cable. |
3300 |
2018-01-19 |
Lab none |
Shipped |
Factory |
This CR6 became unresponsive for no know reason - only a green light on the RJ45 jack. It was sent to Campbell on RMA 27680 for evaluation and repair. |
3299 |
2018-01-15 |
Lab none |
Damaged |
Lab none |
I tried to install this CR6 at Tonzi for the new co2 profile on the Tower but it was non-functional. It tested fine in the lab this morning, but was a brick in the field. The green light on the RJ45 jack comes on but no other indication that it has power. |
3172 |
2017-10-19 |
East End-Temp1 Data Logger Met |
Removed |
Lab none |
Removed with decommissioning of East End Open Water tower (temporary tower 1) |
2855 |
2017-06-01 |
Lab none |
Installed |
East End-Temp1 Data Logger Met |
Installed at the East End Temporary Tower #1 Open Water location with the initial set up of this site. |
1963 |
2015-08-21 |
Factory |
Calibration |
Factory |
Initial factory calibration. Recommended interval: 3 years of service. |