Equipment and Calibration History
History for Campbell Sci. AM25T 4852
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
3092 | 2017-09-26 | East End-Temp2 Multiplexer Thermocouples | Removed | Lab none | Brought back to the lab with the decommissioning of the East End closed canopy temporary tower 2 |
2864 | 2017-06-14 | Lab none | Installed | East End-Temp2 Multiplexer Thermocouples | Installed on the East End Temporary Tower #2 Closed Canpoy location with the initial tower set up. |
2631 | 2017-02-24 | Factory | Returned | Lab none | Returned from Campbell after repair for: 12V screw terminal damage, internal thermistor damage. Campbell says: Able to verify customer problem. Replaced terminal connector (P22C) 12V connector. Replaced internal thermistor (Z16) due to leads being broken off. Bench tested unit, calibrated unit with documentation, final tested to verify correct operation. |
2632 | 2017-02-21 | Factory | Calibration | Factory | Calibrated by Campbell. Good for two years. |
1939 | 2015-09-08 | Lab none | Maintenance | Lab none | I took this multiplexer apart and cleaned it. Some of the screw terminals were tough to turn - I sprayed them with WD40. There was a little bit of corrosion on and near the 12V terminal. |
1850 | 2015-05-12 | Sherman Island Multiplexer Thermocouple | Removed | Lab none | Removed with the decommissioning of the Sherman Cow Pasture site. |
37 | 2010-02-05 | Sherman Island Multiplexer Thermocouple | Location | Sherman Island Multiplexer Thermocouple | Currently connected to the met CR10X sn47253 at Sherman. mostly soil temeraptures. |