2583 | 2017-01-17 | Lab none | Installed | East End Anemometer | Installed on East End tower to replace sn 153804. |
2573 | 2017-01-06 | Lab none | Upgrade | Lab none | Upgraded firmware to correct for W bug.
Old SN: Y124215 New SN: X124215 Upgrade file: Y074417-Y124215-m.gcu, Y012345-s.gcu, X120908 X124215-m.gcu Old firmware: 2329-114 New firmware: 2329-116-01 All Checksum fields report PASS. There are multiple upgrade .gcu files. During the initial upgrade using file "Y074417-Y124215-m.gcu," the sensor was set at 57,600 baud rate. This baud rate caused an error in the upgrade, so the SN reset to Y012345. Gill sent me the second upgrade file, "Y-12345-s.gcu," which reset the sensor to SN X124215. Finally, I used the third upgrade file, "X120908 X124215-m.gcu," to upgrade the firmware. Log file: WMUpgradeLog_X124215_170106_123318.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. ------- THE FOLLOWING FIELDS HAVE CHANGED Baud Rate : 19200 Baud Rate : 57600 Output rate : P1 1 Hz Output rate : P8 20 Hz Output Channel 3 : W output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER======================== Serial Number : X124215 ---- SW VERSION -------------------------------------------- Anemometer Firmware Version: 2329-116 ---- HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Baud Rate : 57600 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P1 1 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : SOS Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound |
2547 | 2016-12-19 | Lab none | Maintenance | Lab none | This was outputting sonic T instead of SOS. I changed its settings to output SOS. Everything else looked okay. Its firmware is 2329-114 - so far not update tool to fix w bug for this version. |
2485 | 2016-10-06 | Bouldin Alfalfa Anemometer | Removed | Lab none | Removed from the Bouldin Alfalfa tower. This sensor was installed to test the system. Replaced by sn 153806. |
2437 | 2016-08-31 | Lab none | Installed | Bouldin Alfalfa Anemometer | Installed on the Bouldin Alfalfa tower to replace sn 153806 that has a spiking problem. |
2145 | 2016-01-12 | East End Anemometer | Removed | Lab none | Removed from tower for calibration. The SOS was reading unexpectedly low. |
1911 | 2015-07-20 | Lab none | Installed | East End Anemometer | Installed on the East End Tower to upgrade the eddy system - replacing older equipment |
1890 | 2015-06-09 | Loaned | Returned | Lab none | Stephen Chan returned this sonic that he had borrowed to test the LI-7550 analog inputs. |
1889 | 2015-02-13 | Lab none | Loaned | Loaned | Loaned to Stephen Chan at Ameriflux at LBNL for testing of LI-7550 analog channels. |
1617 | 2014-10-09 | Mayberry-Portable | Removed | Lab none | Removed with the rest of the portable tower equipment. End of the portable tower campaign. |
1288 | 2014-04-28 | Lab none | Installed | Mayberry-Portable | Installed on the Mayberry portable tower. |