Equipment and Calibration

View History

Equipment Entry

Instrument:Anemometer
Manufacture:Gill
Model No:WM 1590
Serial No:120906
Tag No:NULL
Acquisition Date:2012-03-13
Project:DWR-Wetland
Location on 2021-09-02:Gilbert Tract, NULL
Notes: 4x14bit analog out, 20Hz, Firmware version 2329-116-01

Documents

PackingSlip: ps_Dynamax_20120313.pdf
Report: SW Soinc Compare - 2021-05-26

History

Rec No.DateLoc FromTitleLoc ToNotes
5493 2021-09-02 Sherman Wetland Anemometer Moved Gilbert Tract Anemometer Moved from SW to GT with the decomissioning of SW and the beginning of GT
2605 2017-01-31 Lab none Installed Sherman Wetland Anemometer Installed at Sherman Wetland tower to replace sonic sn 153803. The old sonic does not have the new firmware fixing the w-bug, this new sonic has the new firmware.
2599 2017-01-25 Lab none Upgrade Lab none Upgraded firmware to correct for W bug.

Old SN: Y120906
New SN: X120906
Upgrade file: Y074417-Y124215-m.gcu
Old firmware: 2329-114-01
New firmware: 2329-116-01

All Checksum fields report PASS.

Log file: WMUpgradeLog_Y120906_170125_153219.txt

Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting.

-------

THE FOLLOWING FIELDS HAVE CHANGED

Baud Rate : 19200
Baud Rate : 57600

THE NEW CONFIGURATION IN FULL

==== DETAILS OF DETECTED ANEMOMETER========================

Serial Number : X120906

---- SW VERSION--------------------------------------------
Anemometer Firmware Version: 2329-116

---- HW DETAILS OF ANEMOMETER------------------------------
Detected Anemometer Class : WindMaster 20Mhz
Number of AXIS on this unit : 3
DAC resolution of attached unit : 14
Number of DAC channels : 4
Maximum sampling rate : (8) P8 20 Hz

---- USER CONFIGURATION OF ANEMOMETER----------------------
Baud Rate : 19200
Baud Rate : 57600
Output Message Format : M1 UV(W)
Output Message Velocity Units : m/s
Output Message Terminator :
Retries : ON
Instantaneous Sampling : OFF
Factory Calibration : ON
Physical Communications : Auto
Resolution settings : High resolution
Alignment settings : North to spar
Averaging settings : Averaging inactive
Output rate : P8 20 Hz
Polar angle minimum magnitude: 0.050 m/s
Power-On Message : Display Power On Message
ASCII Format : Comma Separated (CSV)
SOS and Sonic Temperature display : SOS
Analogue output mode : -5V to 5V
Unit Identifier : 'Q'
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 3 : W output (full range = 5m/s)
Output Channel 4 : Speed of Sound

2584 2017-01-19 Twitchell Rice Anemometer Removed Lab none Removed from Twitchell Rice tower and returned to lab for firmware upgrade. It was replaced by sn 153805.
1496 2012-05-07 Lab none Installed Twitchell Rice Anemometer Installed at the rice to replace sn 000024