History for Sensorex DO6400/T DO-07
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
1196 | 2014-01-28 | Lab none | Disposed | Lab none | This probe is no longer working - probably due to corrosion. |
1172 | 2014-01-10 | Mayberry DO Sensor B | Removed | Lab none | The probe was reading high and seemed corroded so it was removed to be serviced in the lab. |
956 | 2013-03-14 | Lab none | Installed | Mayberry DO Sensor B | Reinstalled lower DO probe at the Tule tower |
944 | 2013-03-12 | Lab none | Calibration | Lab none | DO-07 was calibrated in the lab: M = 8.73/36.8 (mg/L)/mV (aka ppm/mV) @ temp = 21.5 C and press.= 101.5 Kpa |
942 | 2013-03-12 | Lab none | Maintenance | Lab none | Replaced membrane. Previous one appeared to be torn. readings seem much better now. |
943 | 2013-02-28 | Lab none | Removed | Lab none | Removed for testing in lab since when holding it out of water at Mayberry the readings were decreasing instead of going back to atmospheric. |
940 | 2013-02-28 | Mayberry DO Sensor B | Removed | Lab none | Removed for testing in lab since when holding it out of water at Mayberry the readings were decreasing instead of going back to atmospheric. |
804 | 2012-09-13 | Lab none | Installed | Mayberry DO Sensor B | DO-07 was installed in the deep open water position at Mayberry. It is about 45cm below the surface. |