Instrument:Heat Flux Plate
Manufacture:Hukseflux
Model No:HFP01
Serial No:001603
Tag No:
Acquisition Date:2006-01-01
Project:
Location on 2022-02-09:Vaira,
Notes:

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
5683 2022-02-09 Vaira Heat Flux Plate A Location Vaira Heat Flux Plate A Wired as SHF #1
1328 2012-11-07 Lab none Installed Vaira Reinstalled after calibration by Sebastian in BC Canada. The original calibration factor was found to be still correct. It was reinstalled in a different location near another SHF plate to the north and cleared the litter from the new location.
775 2012-09-05 Vaira Removed Lab none Removed from the Vaira tower for calibration. This sensors was burried under the end of the PVC conduit and under a gopher mound at about 10cm below the surface. This was the first of the heat flux plates.
5698 2006-10-28 Lab none Installed Vaira Heat Flux Plate A Installed to replace existing plate that had failed months before. This entry was pieced together from notes and examination of the data.
126 2005-09-29 Factory Calibration Factory Initial factory calibration 60.3 uV/W/m2