History for Apogee SI-121 1282

Rec No.DateLoc FromTitleLoc ToNotes
1789 2015-04-20 Factory Disposed Disposed Apogee says: Upon inspection it appears that the thermopile detector has some marks on the FOV glass. This is most likely what is causing the sensors to read so far off from your reference point. I can attempt to calibrate out those marks on the sensor but I am unsure if the uncertainty will be within the 0.1°C specification. The other option is for you to trade in these units for new one’s at a substantial 40% discount, bringing the price to 3 per sensor.

We decided to trade the sensor in for a new one.
1788 2015-04-13 Lab none Shipped Factory A laboratory check showed bad readings from the target thermopile. The readings fluctuated wildly and sometimes out of range. Sent to Apogee on RMA 21085
1758 2015-03-24 Mayberry Infrared Temperature Water Removed Lab none Removed from the Mayberry tower for use by Ana at Tonzi. This sensor was pointing at the water.
347 2011-03-01 Lab none Installed Mayberry Infrared Temperature Water Installed on the Mayberry Tule Tower facing open water.
3811 2010-12-09 Tonzi Removed Lab none Removed from Tonzi with the end of the surface temperature experiment.