Equipment and Calibration

View History

Equipment Entry

Instrument:Pyranometer
Manufacture:Kipp & Zonen
Model No:CMP 11
Serial No:996981
Tag No:
Acquisition Date:1999-09-07
Project:DOE-TCP
Location on 2023-08-03:Lab, 23m
Notes: Initial factory calibration: 5.03 uV/(W/m2)

Documents

Calibration: cal_KippZonen_CM11_sn996981_19990907.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
6355 2023-08-03 Lab none Repair Lab none I took this pyranometer apart and soldered a new short cable to the contacts on the bottom side of the sensor. The wiring colors are the same: red+, blue-, shield. There was a white wire that doesn't seem to be used which I replaced with a yellow so that it is accessible outside the case. It needs new desiccant and probably a longer wire spliced on to be used on a tower. There is a scratch on the outer dome and it needs an inter-comparison.
3001 2017-08-25 Tonzi-Tower Removed Lab none Removed from Tonzi Tower and brought back to the lab. It is a redundant measurement with the 4-way radiometer. Cables are decayed and need to be replaced.
325 2001-05-20 Lab none Installed Tonzi-Tower Installed on the top of Tonzi Tower. [Gleaned from notes]