Equipment and Calibration
View History
Equipment Entry
|
Documents
Calibration: cal_KippZonen_NRLite_sn990352_20021010.pdfCalibration: cal_KippZonen_NRLite_sn990352_19990729.pdf
Report: Oxford Tract Radiometer Comparison - 06/2019 - 2019-06-18
Report: Oxford Tract Radiometer Comparison - 09/2018 - 2019-09-16
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
4194 | 2019-10-01 | Lab none | Calibration | Lab none | New calibration factor of 14.5 uV/(W/m2).
Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 09/2018" for details. https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=27 I'm not sure why the calibration factor is 5% different from the previous calibration in the summer, and which is the "better" calibration to use. The summer had a wider range of Rnet values, but the sensors were more clean in the fall. I think the fall calibration is better. |
4052 | 2019-08-16 | Lab none | Calibration | Lab none | New calibration factor of 13.7 uV/(W/m2).
Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. |
3170 | 2017-10-19 | East End-Temp1 Net Radiometer | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
2915 | 2017-07-03 | East End-Temp1 Net Radiometer | Location | East End-Temp1 Net Radiometer | Checked serial number. |
2892 | 2017-06-01 | Lab none | Installed | East End-Temp1 Net Radiometer | Installed at the East End Temporary Tower #1 Open Water location with the initial set up of this site. |
2708 | 2017-04-04 | Twitchell Rice Net Radiometer | Removed | Lab none | Removed with Twitchell Rice tower decommissioning. |
1739 | 2009-04-03 | Lab none | Installed | Twitchell Rice Net Radiometer | Installed with the initial set up of the Twitchell Rice tower. |
2331 | 2004-05-01 | Tonzi-Floor | Removed | Lab none | Remove net radiometer from met station near tower base - unit to the east of the tower |
2330 | 2002-12-06 | Lab none | Installed | Tonzi-Floor | Installed on CR23x3 with the soil moisture probes at Tonzi. On the floor to the east of the tower. |
4050 | 1999-07-29 | Factory | Calibration | Lab none | Initial factory calibration 14.2uV/(W/m2) |