Rec No. | Date | Loc From | Title | Loc To | Notes |
454 |
2011-12-01 |
Lab none |
Installed |
Tonzi Heat Flux Plate E |
Reinstalled in the field in its orginial location. The ground was dry and crumbly. |
453 |
2011-11-30 |
Lab none |
Repair |
Lab none |
I removed the cable that had cracking and rotting inslation and replaced it with new shielded, two conductor, PVC jacketed cable. New red lead to old green and new black lead to old white. |
452 |
2011-11-16 |
Tonzi Heat Flux Plate E |
Removed |
Lab none |
I removed this heat flux plate from the Tonzi understory location due to bad readings. I found that a gopher had cut the cable. In addition the the insulation on the cable was cracked and rotting. |
5988 |
2001-01-05 |
Lab none |
Installed |
Tonzi Heat Flux Plate E |
Likely first installation date and location estimated from notes and data. |
139 |
1999-09-09 |
Factory |
Calibration |
Factory |
Initial factory calibration: 59.5 uV/W/m2 |