Equipment and Calibration History

History for Hukseflux HFP01 000144

Rec No.DateLoc FromTitleLoc ToNotes
454 2011-12-01 Lab none Installed Tonzi Heat Flux Plate E Reinstalled in the field in its orginial location. The ground was dry and crumbly.
453 2011-11-30 Lab none Repair Lab none I removed the cable that had cracking and rotting inslation and replaced it with new shielded, two conductor, PVC jacketed cable. New red lead to old green and new black lead to old white.
452 2011-11-16 Tonzi Heat Flux Plate E Removed Lab none I removed this heat flux plate from the Tonzi understory location due to bad readings. I found that a gopher had cut the cable. In addition the the insulation on the cable was cracked and rotting.
5988 2001-01-05 Lab none Installed Tonzi Heat Flux Plate E Likely first installation date and location estimated from notes and data.
139 1999-09-09 Factory Calibration Factory Initial factory calibration: 59.5 uV/W/m2