Equipment history at Tonzi| ALL | | Anemometer || Barometer || Battery Bank || Camera || CO2 Probe || Data Logger || Forced Diffusion || Gas Analyzer || Heat Flux Plate || Infrared Temperature || Interface || LED sensor || Moisture Point || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || Precipitation || Radio || RH & Temp || Soil Moisture Probe || Soil Water Potential || Solar Controller || Solar Panel || Water Level || Weather Station |
SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date
Date | Serial No. | Model | Action | Ref | Notes |
---|---|---|---|---|---|
Precipitation | |||||
2018-10-17 | 24872-999 | TR-525I-R1 | Calibration | Got 88 of 100 tips - adjusted screws 2.5 turns CW | |
2019-10-24 | 24872-999 | TR-525I-R1 | Calibration | I connected the bucket to a CR23x dripping 475ml of water through the bucket which should result in 100tips. I got 100 tips. | |
2021-10-06 | 24872-999 | TR-525I-R1 | Calibration | none | Tested with our new method using the peristaltic pump and 237ml of water which should result in 50 tips. 1/2 turn of the screws (CW+, CCW-) should adjust the number of tips by 2-3% (1-1.5 tips). Got 54 tips – I turned the screws two turns CCW |
2012-01-09 | 48102-611 | TE525 | Installed | Installed on the ground at SP3 southeast of the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ. | |
2019-11-07 | 48102-611 | TE525 | Calibration | Plug tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket which should result in 100tips. Got 103 tips and did not adjust the bucket. | |
2021-10-06 | 48102-611 | TE525 | Calibration | none | Tested with our new method using the peristaltic pump and 237ml of water which should result in 50 tips. 1/2 turn of the screws (CW+, CCW-) should adjust the number of tips by 2-3% (1-1.5 tips). Got 50 tips – No adjustment |
2012-01-09 | 48130-611 | TE525 | Installed | Installed on the ground at SP1 inside the soil co2 fence. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ. | |
2018-10-17 | 48130-611 | TE525 | Calibration | Got 88 of 100 tips - adjusted screws 2.5 turns CW | |
2019-10-24 | 48130-611 | TE525 | Calibration | I connected the bucket to a CR23x dripping 475ml of water through the bucket which should result in 100tips. I got 104 tips. I did not adjust it. | |
2020-12-18 | 48130-611 | TE525 | Removed | none | Removed from OSU SP1 for experiments in the lab. The cable was disconnected inside the tipping bucket and left in place in the field. |
2012-01-09 | 48132-611 | TE525 | Installed | Installed on the ground at SP2 near the Tower. This is the initial setup of the site by Richard Cuenca and Yutaka Hagimoto of Oregon State Univ. | |
2018-10-17 | 48132-611 | TE525 | Calibration | Got 134 of 100 tips - adjusted screws 4 turns CCW | |
2019-11-07 | 48132-611 | TE525 | Calibration | Plug tipping bucket into the calibration CR23x and dripped 475ml of water through the bucket which should result in 100tips. Got 103 tips and did not adjust the bucket. | |
2021-10-06 | 48132-611 | TE525 | Calibration | none | Tested with our new method using the peristaltic pump and 237ml of water which should result in 50 tips. 1/2 turn of the screws (CW+, CCW-) should adjust the number of tips by 2-3% (1-1.5 tips). Got 52 tips – I turned the screws one turn CCW. |