Tonzi: 2005-08-04 10:14:00 (DOY 216)
Author: Ted Hehn
Others:

When: Wed 4-6Aug05, JD-215-217
Where: Tonzi
Who: Ted Hehn
Weather Hot & clear

Trip Purpose: Repair CR23x3 at Vaira and Tower CO2 profile pump failure

Tonzi Ranch:
4Aug 10:14 arrive at tower and profile system pump is not running. 12 volts present on pump power leads 2” from pump. Cut positive lead and insert DMM. No current is being drawn by pump motor.

Begin diagnostics on profile system by removing DC/DC converter board from system and discover what appears to be cold or failed solder joint on negative lead of pump/converter output pin. Repair and reassemble system. Check continuity on all leads to and from converter board. No other problems discovered.

Power up CO2 profile system and run zero/span sequences. All appears well.

Remove DCS900 webcam and install Vivotek IP6112 webcam, bat and script files.
Remove DCS900 webcam software shortcut from startup menu.

Note: I changed the power wire configuration for the LI7500/Sonic/Webcam and it now appears that the level of noise visible on the H2O and CO2 traces has been SIGNIFICANTLY reduced!!! When I return with the next round of data we should run a power spectra and demonstrate whether or not this is true???

Create a “Scheduled Task” in control panel to actuate Batch file IP6112.bat beginning every day at 7:00, once per hour for a 12 hour duration.

5Aug05 14:30 arrive at tower and pump is not running. Cold solder joint not the problem. Strike the pump with screwdriver and it starts running. Remove 20 Ga power cable between bulkhead and pump and replace with 16Ga conductor.

6Aug05 10:00 Arrive at tower and pump was not running. Shut down system and remove DC/DC converter board and replace pump DC/DC converter. Restart system and perform multiple Zero/Span sequences. OK all appears well.

Install wheels on pull cart.

Field Data

No tank data

No TZ Well data found

No TZ Moisture Point data found

No VR Moisture Point data found

No TZ grass heigth found

No VR grass heigth found

Tree DBH

DateTag#Reading (mm)

No water potential data found

No TZ Grass Biomass data found

No TZ OakLeaves Biomass data found

No TZ PineNeedles Biomass data found

0 photos found

Graphs display:
16 sets found

No data found for TZ_Fenergy.
11 TZ_tmet graphs found

Tonzi Tower Met: Temperature and Power

Tonzi Tower Met: Solar Radiation

Tonzi Tower Met: IRT Sensors

LED Sensor Incoming

LED Sensor Outgoing

4-Band Solar Radiation

Air Temperature Profile

Relativity Humidity Profile

Tonzi Tower PRI

SRS PRI/NDVI

SRS NDVI

No data found for TZ_soilco2.
Can't check TZ_mixed
3 TZ_mixed graphs found

Tonzi PPT

Tonzi CO2 and H2O Densities

Air Pressure

No data found for TZ_fflux.
8 TZ_fmet graphs found

Soil Heat Flux

Tonzi Soil Temperature Set A

Tonzi Soil Temperatures Set B

Tonzi Soil Temperatures Set C

Tonzi Soil Temperature Set D

Well Level

Well Water Temperature

Soil Water Potential

No data found for TZ_tram.
No data found for TZ_co2prof.
No data found for TZ_irt.
1 TZ_windprof graphs found

Wind Speed Profile

No data found for TZ_tflux.
2 TZ_sm graphs found

Floor Solar Radiations

TZ Soil Moisture

Can't check TZ_met
1 TZ_met graphs found

Floor and Tower RH

Can't check TZVR_mixed
No data found for TZ_osu.
No data found for TZ_fd.
Tonzi Z-cam data

plot of Tonzi z-cam LAI

plot of Tonzi z-cam gap fration

 
 

Login
Select Notes:

This material is based upon work supported by the National Science Foundation and US Department of Energy. Any opinions, findings, conclusions, or recommendations expressed in the material are those of the author(s) and do not necessarily reflect the views of the supporters.