View history entry 1418 for Vaisala GMP221 sn: H0960006

Date: 2012-06-27
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Trueppm = Probeppm*1.003 + 133; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 3056ppm, and 10000ppm. Ta=25.9C, P=101.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara.