Instrument:DO Sensor
Manufacture:Sensorex
Model No:DO6400/T
Serial No:DO-05
Tag No:
Acquisition Date:2012-03-12
Project:DWR-Wetland
Location on 2012-09-06:West Pond, -15cmwater
Notes: Campbell part number CS511

Documents

PackingSlip: ps_Campbell_20120312.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
800 2012-09-06 Lab none Installed West Pond DO Sensor Upper DO-05 was resintalled in the shallow location at West Pond. The lab check went fine and it has been decided that the low values are correct and that the water is just very oxygen poor.
799 2012-08-30 West Pond DO Sensor Upper Removed Lab none DO-05 was removed from the water at West Pond due to continued low readings. The sensor will be checked in the lab and reinstalled.
798 2012-08-16 West Pond DO Sensor Upper Serviced West Pond DO Sensor Upper DO-05 was serviced in the field due to continued unexpectedly low readings. The inside was rinsed with vinegar and scrubbed with a tooth brush. The electrolyte was replaced.
797 2012-08-02 West Pond DO Sensor Upper Serviced West Pond DO Sensor Upper DO-05 was serviced in the field due to unexpectedly low readings. The electrolyte was replaced.
538 2012-04-17 Lab none Installed West Pond DO Sensor Upper Installed on post northwest of scaffolding at west pilot pond. Not running because of forgotten power connector.
808 2012-04-16 Lab none Calibration Lab none DO-05 was calibrated in the lab: M = 8.56/34.49053 in (mg/L)/mV (aka ppm/mV) (@ 23C) @ 5 cm below water
 
 