Equipment and Calibration
View History
Equipment Entry
|
Documents
No documents found
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6191 | 2023-02-15 | Lab none | Returned | Returned none | Returned to Adina during our Elkhorn Slough visit |
6195 | 2023-01-25 | Gilbert Tract none | Removed | Lab none | USGS removed this EXO from the Gilbert Tract channel and stashed it on the bank. Joe retrieved it from the site at the end of the day and brought it back to the lab. |
6062 | 2022-11-03 | Gilbert Tract | Calibration | Gilbert Tract | fDOM (QSU and RFU), Last cal: 10/31/2022 4:14:35 PM EXO sn 22C100485 EXO/Sensor Firmware: 1.0.86/3.0.0 UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,0.51 QSU,0.01 QSU,0.00 QSU,16.967 °C Pt2,316.50 QSU,219.51 QSU,316.49 QSU,0.00 QSU,10.855 °C ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,0.12 RFU,0.00 RFU,0.00 RFU,15.878 °C Pt2,105.50 RFU,74.86 RFU,105.50 RFU,0.00 RFU,11.257 °C |