Equipment and Calibration

View History

Equipment Entry

Instrument:Gas Analyzer
Manufacture:Licor
Model No:LI-7500RS
Serial No:75H-0128
Tag No:
Acquisition Date:2022-04-22
Project:Borrowed
Location on 2022-04-25:Lab,
Notes: Borrowed from Ameriflux for 2022 advection setup. Came with an IRGA cable. Due back 11/1/2022.

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
6134 2022-11-15 Lab Calibration Lab Last calibration: 2021-09-21
AGC=102 SDM=0 DAC1=- DAC2=-
Ambient P=101.1kPa, Tair=23.3C, Tdew=0.53C
Zero Air at 1800psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00
CO2 zero 1.1643 -> 1.1626 Offset: -3.68ppm
CO2 span 0.9978 -> 1.0087 Offset: -7.23ppm
H2O zero 1.0603 -> 1.0599 Offset: -2.2ppm
H2O span 1.0172 -> 1.0374 Offset: -0.44ppm
Notes: Calibrated on box sn AIU-2312. This pair is kept together because they are an RS model and belong to AMP.
6067 2022-11-02 Bouldin Alfalfa-ADEX Gas Analyzer Bouncer CO2/H2O Removed Lab none Removed from top of the Bouncer tower at 3.96m alg at the end of the 2022 advection experiment at BA
5852 2022-04-25 Lab none Installed Bouldin Alfalfa-ADEX Gas Analyzer Bouncer CO2/H2O Installed on top of the Bouncer tower at 3.96m alg for the advection experiment at BA
6136 2021-09-20 Factory none Calibration Factory none Last calibration:
AGC= SDM= DAC1= DAC2=
Ambient P=kPa, Tair=C, Tdew=C
at psi, 750 at Tdewspan=12
CO2 zero -> 1.1626 Offset: ppm
CO2 span -> 1.0087 Offset: ppm
H2O zero -> 1.0599 Offset: ppm
H2O span -> 1.0374 Offset: ppm
Notes: Full calibration values on yellow slips