Equipment and Calibration History
History for Vaisala HMP45AC U3030032
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
2750 | 2017-05-05 | Factory | Disposed | Disposed | Not repairable, so disposed of at CSI. |
2749 | 2017-04-03 | Lab none | Serviced | Factory | Shipped to Campbell Sci for calibration. CSI Technician notes: Tested & inspected. 2x4 connectors have corrosion, need replaced. Needs new o-rings. While testing sensor head, also found temp chip out of spec, drifting high in high humidity. Temperature chips are no longer available for this sensor. Sensor can not be repaired. Dispose of at CSI. |
1067 | 2013-08-28 | Twitchell Rice RH & Temp | Removed | Lab none | Removed from the Twitchell Rice tower due to bad looking RH readings. Need to be tested in the lab. Replaced by sn U3030046 |
265 | 2011-01-18 | Lab none | Installed | Twitchell Rice RH & Temp | This sensor was checked in the lab against the Vaisala hand-held and was very close +-1C +-1%. It was then installed on the Twitchell tower to replace the one there that didn't seem to be working. |
95 | 2010-03-18 | Sherman Island | Removed | Lab none | Removed from the soil moisture CR10X at Sherman Island. Values were bad and duplicate measurment are on the tower. Also needed the solar radiation shield for Vaira. |
56 | 2010-02-09 | Sherman Island | Location | Sherman Island | This is currently connected to the CR10X collection soil moisture from the echo probes. |
96 | 2007-01-02 | Factory | Calibration | Factory | Calibrated at Campbell: Tested and inspected, replaced O-rings, calibrated |
4894 | 2003-01-21 | Tonzi none | Removed | Lab none | Removed from Tonzi. Replaced by sn W0510026 |