Equipment and Calibration History
History for Vaisala GMP222 X3830001
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
592 | 2012-05-02 | Factory | Disposed | Disposed | Filter was cracked, probe completely replace - new probe is serial number H0630070 |
575 | 2012-03-28 | Lab none | Shipped | Factory | Sent to Vaisala for repair. Reporting errors 10, 11 and 12 readings unreasonable. |
1367 | 2010-03-02 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.065 + -1.94; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 383ppm, 597ppm, and 2238ppm. Ta=22.2C, P=101.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Jaclyn. |
4631 | 2010-01-08 | Tonzi CO2 Probe Tree1 Profile -0.02 | Location | Tonzi CO2 Probe Tree1 Profile -0.02 | Pulled out for ambient air reading and put back. |
4541 | 2008-03-06 | Lab none | Installed | Tonzi CO2 Probe Open Profile -0.02 | Installed in the East open grass profile at -2cm. mV and ppm readings in air were recorded before installation. |