Equipment and Calibration History
History for Kipp & Zonen PAR-LITE 010159
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
3764 | 2018-10-25 | Lab none | Installed | Mayberry PAR Sensor Incoming | Installed at MB in the incoming position to replace sn 10162, which was brought back to the lab for repair/calibration. |
4533 | 2018-09-28 | Lab none | Calibration | Lab none | Calibrated in the lab by sticking it out the south-facing window in Hilgard 205 with the reference sensor sn 161073. The factory calibration was 5.16 µV/(µmol/m2·s) and the new calibration was 5.59 µV/(µmol/m2·s). For details see the RT report "PAR comparison 09-2018" https://nature.berkeley.edu/biometlab/bmetdata/rtreport.php?screen=view&id=14 |
3716 | 2018-09-24 | Lab none | Repair | Lab none | Trimmed old cable, soldered on ~10m cable, and secured cable-sensor junction with small L-bracket and zip-ties. |
3004 | 2017-08-25 | Tonzi-Tower | Removed | Lab none | Removed from Tonzi Tower. This sensor has been reading high for a long time. The cable is decayed and the sensor needs to be re-calibrated. This sensor was entered into the database. |
5159 | 2004-04-30 | Lab none | Installed | Tonzi PAR Sensor Incoming Cal | Installed on the radiation boom in the incoming position for inter-comparison with the existing Licor Quantum sensor. |
4535 | 1999-02-09 | Factory none | Calibration | Factory none | Initial factory calibration 5.16 uV/(umol/m2/s). |