Equipment and Calibration History
History for Forerunner-Eosense GP-001 GP20140016
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
4860 | 2020-10-08 | Lab none | Shipped | Factory none | Sent to Eosense for inspection |
3795 | 2018-11-06 | Lab none | Tested | Lab none | Connected to a CR6 and recorded 10s data. The sensor was maxing out (reading ~5V) on both 0-2% and 0-6% ranges. The connection between sensor and cable seemed secure and clean. The sensor should go back to the factory. |
3751 | 2018-10-18 | Sherman Wetland CO2 Probe | Removed | Lab none | Removed from Sherman Wetland due to railed readings not responding the atmospheric CO2. |
3041 | 2017-08-16 | Lab none | Installed | Sherman Wetland CO2 Probe | Installed at SW. I waterproofed the cable-sensor connection with multiple layers of liquid electrical tape and silicone caulk. |