Equipment and Calibration
View History
Equipment Entry
Instrument: | CO2 Probe |
Manufacture: | Forerunner-Eosense |
Model No: | GP-001 |
Serial No: | GP20140084 |
Tag No: | |
Acquisition Date: | 2015-12-07 |
Project: | CDFW |
Location on 2021-01-21: | Lab, |
Notes: |
0-2% probe. Sensor dead. |
|
|
Documents
No documents found
Report: Eosense CO2 Probe Compare - 2018-04-23
Report: Eosense CO2 Probe Compare - 2018-09-07
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
5054 |
2021-01-21 |
Lab none |
Tested |
Lab none |
This sensor is dead. Tested in the lab and no response from CO2 or T. The connector at the top of the sensor twists loose from the body of the sensor. There was some corrosion in the connector head and the cable. |
5052 |
2021-01-17 |
Sherman Wetland CO2 Probe |
Removed |
Lab none |
Removed from SW and brought back to the lab b/c the sensor is bad and only outputting noise |
3784 |
2018-11-01 |
Lab none |
Installed |
Sherman Wetland CO2 Probe |
Installed at SW to replace sn GP20140016, which was removed b/c it was maxing out. |
3353 |
2018-03-07 |
Twitchell Alfalfa |
Removed |
Lab none |
Removed from the -2cm location at Twitchell Alfalfa because the readings have been funky for a while. Probably will be replaced at a later date. |
2256 |
2016-04-28 |
Lab none |
Installed |
Twitchell Alfalfa |
Replaced probe sn GP20140023 at 2cm depth since that was reading too high. |