Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
2535 | 2016-11-17 | Lab none | Installed | Vaira Pond | Installed at the main pond at Varia. This will collect water level data. |
2524 | 2016-11-15 | Lab none | Maintenance | Lab none | The internal lithium battery was dead. I replaced the battery. |
377 | 2011-06-18 | Lab none | Loaned | Siyan | Loaned to Siyan for her work at Teakettle Forest. |
371 | 2011-06-01 | Factory | Returned | Lab none | Returned from repair at Campbell. Campbell says, "Unit would not respond when received. Replaced PLD chip and installed OS Rev 23. Tested unit on the bench. Ran unit in the environmental chamber. Retested unit and checked calibration. Verified unit would work with SDI12 sensors."
Calibration good until May 25, 2013. |
361 | 2011-05-17 | Lab none | Repair | Factory | Sent to Campbell Scientific for evaluation and repair on RMA 13443 |
358 | 2011-05-13 | Sherman Island Data Logger Fast Rad | Removed | Lab none | Removed from Sherman diffuse radiation because I could not connect to it. It seemed to function at night (cool temperatures?) but did not work during the day. |
38 | 2010-02-05 | Sherman Island Data Logger Fast Rad | Location | Sherman Island Data Logger Fast Rad | Currently used to collect the fast radiation data from a CNR1 and diffuse PAR at Sherman. Has wiring panel #26593 |
359 | 2002-07-22 | Factory | Calibration | Factory | Calibrated at Campbell. due 2003. |