Equipment and Calibration History
History for Gill WM 1590 153806
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
5716 | 2022-03-07 | Lab none | Installed | Loaned none | Installed on the urban flux tower on top of Berkeley Way West building |
5386 | 2021-07-20 | Bouldin Alfalfa Anemometer | Removed | Lab none | Removed from Bouldin Alfalfa and replaced by sn 074417 so that the advection experiment could run with four inter-compared sonics. |
2486 | 2016-10-06 | Lab none | Installed | Bouldin Alfalfa Anemometer | Reinstalled on the Bouldin Alfalfa tower replacing sn 124215. This sonic is working correctly. Spikes were caused by the soil moisture probes powering up and browning out the system. |
2438 | 2016-08-31 | Bouldin Alfalfa Anemometer | Removed | Lab none | Removed from the Bouldin Alfalfa tower due to daytime spiking problem likely due to new firmware. Replaced by sn 124215 |
2420 | 2016-08-12 | Lab none | Installed | Bouldin Alfalfa Anemometer | Installed with the initial set up of the site. |
2392 | 2016-08-11 | Lab none | Upgrade | Lab none | Using the Gill provided software tool we upgraded this unit's firmware from version 2329-601-03 to version 2329-701-01. This new firmware is supposed to fix the W bug. |