Rec No. | Date | Loc From | Title | Loc To | Notes |
2802 |
2017-05-31 |
Factory |
Disposed |
Disposed |
Disposed of at Eosense factory because it was not repairable |
2664 |
2017-03-20 |
Lab none |
Shipped |
Factory |
Sent to Eosense for evaluation. Failed after a total 586 days in use. |
2495 |
2016-10-11 |
Mayberry CO2 Probe B |
Removed |
Lab none |
Removed from Mayberry due to odd readings and lowering water for flushing salt |
1898 |
2015-06-25 |
Lab none |
Installed |
Mayberry CO2 Probe B |
Installed in the middle location at Mayberry to replace sn GP20140026 due to corroding connector. This probe has the connector painted to hopefully slow the corrosion. |
1723 |
2015-02-05 |
Mayberry CO2 Probe C |
Removed |
Lab none |
Removed from the bottom position of the stringer at Mayberry. Replaced by sn GP20140023 that has copper mesh over the filter. |
1613 |
2014-10-16 |
Lab none |
Installed |
Mayberry CO2 Probe C |
Installed at 108cm below the surface on the fence post in the open water channel. |