Equipment and Calibration History
History for Forerunner-Eosense GP-001 GP20140025
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
2802 | 2017-05-31 | Factory | Disposed | Disposed | Disposed of at Eosense factory because it was not repairable |
2664 | 2017-03-20 | Lab none | Shipped | Factory | Sent to Eosense for evaluation. Failed after a total 586 days in use. |
2495 | 2016-10-11 | Mayberry CO2 Probe B | Removed | Lab none | Removed from Mayberry due to odd readings and lowering water for flushing salt |
1898 | 2015-06-25 | Lab none | Installed | Mayberry CO2 Probe B | Installed in the middle location at Mayberry to replace sn GP20140026 due to corroding connector. This probe has the connector painted to hopefully slow the corrosion. |
1723 | 2015-02-05 | Mayberry CO2 Probe C | Removed | Lab none | Removed from the bottom position of the stringer at Mayberry. Replaced by sn GP20140023 that has copper mesh over the filter. |
1613 | 2014-10-16 | Lab none | Installed | Mayberry CO2 Probe C | Installed at 108cm below the surface on the fence post in the open water channel. |