Equipment and Calibration History
History for Forerunner-Eosense GP-001 GP20140021
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
2806 | 2017-05-31 | Factory | Disposed | Disposed | Disposed of at Eosense factory because it was not repairable |
2667 | 2017-03-20 | Lab none | Shipped | Factory | Sent to Eosense for evaluation. Bad readings after a total 276 days in use. |
2640 | 2017-02-22 | Lab none | Calibration | Lab none | y = 13.36*x The probe was tested against a new probe with decreasing CO2 from 2.4% to zero. This probe read significantly lower than the standard probe and showed some lag at higher change rates. |
2120 | 2015-12-08 | East End CO2 probe | Removed | Lab none | This 4% CO2 probe was removed from East End. It was replaced by an 8% probe sn GP20150109 |
1902 | 2015-06-25 | Lab none | Installed | East End CO2 probe | Installed this 4% probe at East End due to replaced 2% probe sn GP20140011. |
1672 | 2015-01-20 | West Pond CO2 Probe B | Removed | Lab none | Removed from West Pond. Replaced by sn GP20140022 that now has a copper mesh. |
1580 | 2014-10-02 | Lab none | Installed | West Pond CO2 Probe B | Installed at West Pond to replace the current 2% probe sn GP20140009. |