1554 |
2014-09-11 |
Tonzi-Tower |
Maintenance |
Tonzi-Tower |
I updated the SDI-12 of this sensor from 0 (zero) to 2 (two). This solved the problem of the 23x reading the sensor. The update was done using a CR1000 in SDI-12 transparent mode. |
1553 |
2014-08-21 |
Lab none |
Installed |
Tonzi-Tower |
I installed this sensor on the radiation boom at the tower top. It is facing down collecting reflected radiation. This is an SDI-12 sensor, but the 23x was having trouble reading it. I brought in its twin to troubleshoot in the lab. |