Equipment and Calibration History
History for Vaisala GMP221 K1340007
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
1875 | 2015-06-09 | Twitchell Alfalfa | Removed | Lab none | Removed from Set A at -8cm with upgrade of soil CO2 system |
1862 | 2014-08-13 | Lab none | Installed | Twitchell Alfalfa | Installed near the radiometers at Alfalfa in Set A to the south at -8cm |
2090 | 2014-04-10 | Lab none | Installed | Sherman Island | Installed in Sherman profile A at -8cm to replace sn F4960019 |
1452 | 2014-04-07 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.01 + -11.96; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe. |