Rec No. | Date | Loc From | Title | Loc To | Notes |
6413 |
2023-09-18 |
Lab none |
Loaned |
Loaned none |
There are a sticky note claiming this one was sn 243? Loaned to Gracie Pearsall at UCSC for a informational display at Elkhorn Reserve |
4540 |
2014-12-10 |
Twitchell Rice Heat Flux Plate A |
Removed |
Lab none |
Removed from TW and brought back to the lab for troubleshooting |
1630 |
2014-11-25 |
Lab none |
Installed |
Twitchell Rice Heat Flux Plate A |
Installed in multiplexer on channel 16. The heat flux plate was reading INF, so this needs to be looked at next time—I couldn’t diagnose the problem. |
1631 |
2014-11-21 |
Lab none |
Repair |
Lab none |
There was a break in the insulation and glue near the plate where it had been repaired before. I cleaned up the plate, re-spliced the wire near the plate and sealed the connection with black silicon. It seemed to be functional using a voltmeter and heat. |
1629 |
2014-11-16 |
Twitchell Rice Heat Flux Plate A |
Removed |
Lab none |
SHFP1 was removed from rice since reading INF. Joe will diagnose in lab. |
1341 |
2013-06-07 |
Lab none |
Installed |
Twitchell Rice Heat Flux Plate A |
Installed to replace the one cut off by a tractor (sn 00244). Installed in AM25T channel 16 (SHF1). |
1340 |
2011-10-27 |
Twitchell Rice Heat Flux Plate B |
Removed |
Lab none |
Joe checked the wires on the SHF plates and found that they were rotting away. Brought back to the lab to fix. |
1339 |
2009-04-03 |
Lab none |
Installed |
Twitchell Rice Heat Flux Plate B |
Installed at the rice as SHF2 in channel 17 on the multiplexer. It's calibration coefficient is: 62.1 micrV/(W m-2) |