Equipment and Calibration History
History for Youngryel Ryu LED-4chan RYU-01
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
2702 | 2017-04-05 | Tonzi-Tower | Removed | Lab none | Removed from the Tonzi tower. This sensor has not been giving good readings for awhile. Replaced by one of our LED sensors. |
1176 | 2013-12-15 | Tonzi-Tower | Damaged | Tonzi-Tower | A bird began to poke a hole in the incoming filter and eventually tore the filter completely off. |
1175 | 2013-11-07 | Lab none | Installed | Tonzi-Tower | Reinstalled on the Tonzi Tower. A wiring error swapping the Red and NIR channels was fixed. New Teflon filter. |
1174 | 2013-09-25 | Tonzi-Tower | Removed | Lab none | Removed from the Tonzi tower for repair of missing incoming filter |
1173 | 2013-08-15 | Tonzi-Tower | Damaged | Tonzi-Tower | A bird began to poke a hole in the incoming filter and eventually tore the filter completely off. |