Equipment and Calibration
View History
Equipment Entry
|
Documents
No documents found
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
914 | 2013-01-04 | Lab none | Tested | Disposed | Upon inspection in the lab. this sensor was found to be severely corroded. However, at the time of inspection the sensor was not damp. |
913 | 2013-01-03 | Mayberry CO2 Probe B | Removed | Lab none | Removed from the tules near the tule tower due to bad values. |
659 | 2012-07-26 | Lab none | Installed | Mayberry CO2 Probe B | Installed on the Tule Tower to measure water CO2 concentration. |
1418 | 2012-06-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.003 + 133; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 3056ppm, and 10000ppm. Ta=25.9C, P=101.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |