Rec No. | Date | Loc From | Title | Loc To | Notes |
5966 |
2022-08-24 |
Lab none |
Installed |
Hill Slough none |
Installed with a solar panel and lithium battery at the Hill Slough channel. This datalogger will collect data from water sensors installed in the channel. |
5833 |
2022-05-10 |
Vaira Data Logger Soil CO2 |
Removed |
Lab none |
Removed with the last of the soil co2 system because data not good and to reduce power load. |
5041 |
2020-12-18 |
Vaira Data Logger Soil CO2 |
Location |
Vaira Data Logger Soil CO2 |
Confirmed serial number and location |
5044 |
2020-11-11 |
Lab none |
Installed |
Vaira Data Logger Soil CO2 |
Installed at Vaira in the Soil CO2 system. The 0-5V range on the 23x avoids the problem of needing voltage dividers. |
2990 |
2017-08-17 |
Vaira Data Logger Met |
Removed |
Lab none |
Removed from Vaira where it was collecting met data. It was replaced by a CR1000. |
2465 |
2007-07-24 |
Factory |
Returned |
Lab none |
Returned from repair at Campbell on RMA 4887: "Verified that it would not communicate when received. Unit had several low 12V errors and corrupt intermediate location which caused the communication problems. Most likely caused by lightning strike or repeated low supply problems. Updated Operating System to latest revision 19. calibrated with documentation, temperature tested and given a bench test." |