Instrument:Conductivity Sensor
Manufacture:Campbell Sci.
Model No:CS547A
Serial No:6096
Tag No:
Acquisition Date:2012-03-12
Project:DWR-Wetland
Location on 2024-02-12:Lab, -15cmwater
Notes: Needs interface module module no A547. Calibration constants: Rcable = 20; Kc = 1.411

Documents

PackingSlip: ps_Campbell_20120312.pdf
Report: Conductivity CS547A TempCoef - 2024-02-14

History

Rec No.DateLoc FromTitleLoc ToNotes
6567 2024-02-12 West Pond Conductivity Sensor Removed Lab none Removed from WP with interface sn 3586 and brought back to the lab. Nothing wrong with the data (probably), I just wanted an extra sensor in the lab to play around with TempCoef.
4507 2020-05-29 West Pond Conductivity Sensor Calibration West Pond Conductivity Sensor Calibration in the field with DI water (0 mS) and a YSI solution (10,000 uS). Temperature-corrected mS/cm readings from the data logger, where the temperature correction occurs in the data logger program.
In DI water: 0.00mS/cm, 20.9C
In 10,000 uS calibration soln: 10.3mS/cm, 23.0C
Calibration: (Corrected mScm) = 0.9709 * (Sensor mScm)
540 2012-04-17 Lab none Installed West Pond Conductivity Sensor Installed on post northwest of scaffolding at west pilot pond. Not running because of forgotten power connector.