Equipment and Calibration History
History for Campbell Sci. CS547A 6096
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6567 | 2024-02-12 | West Pond Conductivity Sensor | Removed | Lab none | Removed from WP with interface sn 3586 and brought back to the lab. Nothing wrong with the data (probably), I just wanted an extra sensor in the lab to play around with TempCoef. |
4507 | 2020-05-29 | West Pond Conductivity Sensor | Calibration | West Pond Conductivity Sensor | Calibration in the field with DI water (0 mS) and a YSI solution (10,000 uS). Temperature-corrected mS/cm readings from the data logger, where the temperature correction occurs in the data logger program. In DI water: 0.00mS/cm, 20.9C In 10,000 uS calibration soln: 10.3mS/cm, 23.0C Calibration: (Corrected mScm) = 0.9709 * (Sensor mScm) |
540 | 2012-04-17 | Lab none | Installed | West Pond Conductivity Sensor | Installed on post northwest of scaffolding at west pilot pond. Not running because of forgotten power connector. |