Equipment and Calibration History
History for Kipp & Zonen NR-LITE 990351
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
6532 | 2024-01-11 | Mayberry Net Radiometer Water | Removed | Lab none | Removed from MB tower because it was no longer measuring open water (wetland is filling in with cattails) |
4023 | 2019-07-25 | Lab none | Installed | Mayberry Net Radiometer Water | Installed at MB over the water to replace sn 990350, which was brought back to the lab for repair. |
4051 | 2019-06-18 | Lab none | Calibration | Lab none | New calibration factor of 14.0 uV/(W/m2). Tested this sensor at Oxford Tract against AMP 4-band radiometer CNR4 sn 16084. See RT Report "Oxford Tract Radiometer Comparison - 06/2018" for details. |
2999 | 2017-08-25 | Tonzi-Tower | Removed | Lab none | Removed from Tonzi Tower and brought back to the lab. It is a redundant measurement with the 4-way radiometer. Cables are decayed and need to be replaced. |
3688 | 2001-05-20 | Lab none | Installed | Tonzi | Installed on the Tower top with the initial set up of the site |
4055 | 1999-07-29 | Factory | Calibration | Lab none | Initial factory calibration: 13.9 uV/(W/m2) |