Equipment and Calibration
View History
Equipment Entry
|
Documents
No documents found
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
1816 | 2013-04-05 | Sherman Island | Removed | Lab none | Removed from the box at Sherman. This pump has not been run in the field in a long time and probably needs a check up before it is run again. |
208 | 2010-09-21 | Lab none | Installed | Sherman Island | Installed in the box at Sherman Island replacing Frank's diaphragm pump. Sounds real smooth. |
202 | 2010-09-03 | Factory | Repair | Lab none | Returned from factory rebuild. Major kit with bearings, tip seals, pin crank shafts, and crank shaft. From the factory: "The pumps run very hot at 140torr and it's causing some thermal breakdown of the grease and early bearing failure. What we have done is to pack your bearings with grease around an extra 50% by weight and we think this should get you a longer run life between rebuilds. I think we can expect this pump to make the 12 month mark and maybe longer." |
172 | 2010-08-03 | Lab none | Repair | Factory | The pump was pickup for service at http://provac.com/ |
152 | 2010-07-29 | Lab none | Notes | Lab none | This pump has four serial numbers on it:
End cap: 043071002 Side Plate: DG002 Side Sticker: 044012498 Motor Plate: 84010040551 |
153 | 2010-07-28 | Sherman Island | Removed | Lab none | This pump was removed from the Sherman Island LGR for maintenance. It has 3467 hours on it. |
156 | 2010-03-11 | Lab none | Installed | Sherman Island | This pump was installed in the box at Sherman Island to run the LGR FMA there. The pump had 140 hours on it when it was installed. |
158 | 2010-02-04 | Twitchell Rice | Removed | Lab none | This pump was removed from the box at Twitchell Island where it had been sitting idle. |
155 | 2009-10-22 | Lab none | Installed | Twitchell Rice | This pump was installed in the box at Twitchell Island to run the LGR FGGA there. |